Safety
CAUTION
Risk of injury due to unlocking of the knee joint
After activating the release mechanism / unlocking feature (switch to " 
E-MAG Active system. In order to use the system for walking (switch to "   "), it must first be secured by reactivat­
ing the release mechanism.
CAUTION
Incorrect alignment, assembly or adjustment
Risk of injury due to change in or loss of functionality
► Assembly, adjustment and maintenance operations may only be completed by qualified personnel.
► Observe the alignment, assembly and adjustment instructions.
CAUTION
Unallowable use of lubricants
Risk of injury and loss of functionality as well as damage to the product
► The product is designed to be free of lubricants. Avoid contact with lubricants (e.g. grease, oil).
► After contact with lubricants, promptly clean the product or parts of the product with a degreasing cleaning
agent (e.g. acetone or isopropyl alcohol).
CAUTION
Exposure of the product to unsuitable environmental conditions
Patient injury, damage, brittleness or destruction due to improper handling
► Do not expose the product to condensing ambient humidity or liquids.
► Do not expose the product to abrasive substances (e.g. sand, dust).
► Do not expose the product to temperatures below -10 °C (14 °F) or above +60 °C (140 °F) (e.g. sauna,
excessive sunlight, drying on a radiator).
CAUTION
Improper handling
Risk of falling due to accidentally opening or engaging the lock
► Before loading the orthosis, verify that the switch is in the desired position.
► Avoid accidentally opening or engaging the lock.
► Instruct the patient in the correct usage of the system joints.
CAUTION
Knee not fully extended at heel strike
Risk of injury due to unlocked orthosis joint
► Fully extend the knee prior to heel strike.
NOTICE
Risk of damage to electronic components due to installation errors or improper use
Incorrect handling of the electronics can cause malfunctions and / or damage to the electronics. Please observe
the following safety instructions:
► Please read the information regarding the operation of the electronics thoroughly and instruct the patient in
the use of the electronics (see the information on the functionality of electronic components).
► Only install electronic components in the thigh area.
► All electronic connections should be protected against oxidisation with silicone grease.
► In case of defective electronic components, always replace the entire component with a spare part.
NOTICE
Contact with salt water or water containing chlorine/soap as well as abrasive substances (e. g. sand)
Damage and premature product wear
► Following any contact with the substances identified above, promptly clean the product in accordance with
the section "Cleaning and care".