Guidance and Manufacturer's Declaration – Electromagnetic Immunity
(continued)
The BMC Radiofrequency Puncture Generator is intended for use in the electromagnetic
environment specified below. The customer or the user of the BMC Radiofrequency
Puncture Generator should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Compliance
Electromagnetic environment -
level
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
BMC Radiofrequency Puncture
Generator including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
3 Vrms
Recommended separation
distance:
d
=
. 1 [
3 V/m
=
[
d
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
be less than the compliance level in
each frequency range
Interference may occur in the
vicinity of equipment marked with
the following symbol:
37 DMR RFP-100 3.3 V-13 16-Jan-2012_EN.doc
17
]
P
80 MHz to 800 MHz
. 2
33
]
P
800
MHz to 2.5 GHz
a
should
b