Dimensions:
Weight:
Leakage Current Measurements (No Fault Condition):
Catheter Source Current
Indifferent Electrode Source Current
Catheter Sink Current
Indifferent Electrode Sink Current
Case Leakage (Ground Open, Power Normal)
Case Leakage (Ground Open, Power Reversed)
Dielectric Withstand (Hi-Pot) Test:
Mains to Catheter (4,000 V, 1min)
Mains to Indifferent Electrode (4,000 V, 1min)
Accessory Rated Voltage:
Accessory Device
PowerWire
Guidewire
Nykanen Radiofrequency
Wire
TM
NRG
Transseptal Needle
Toronto Transseptal Catheter
Toronto Transseptal Wire
Width: 12 inches
Length: 14 inches
Height: 6 inches
20 lb. (9.09 kg)
TM
Radiofrequency
30 DMR RFP-100 3.3 V-13 16-Jan-2012_EN.doc
(30 cm)
(35 cm)
(15 cm)
< 10 uA
< 10 uA
< 10 uA
< 10 uA
< 100 uA
< 100 uA
PASS
PASS
Rated Accessory
Voltage (V
)
rms
224
224
224
224
224