FRONT PANEL

5.2.

RF ON/OFF BUTTON/LIGHT

The RF ON/OFF Button/Light is shown in Figure 5-1G.
responsible for the three operational modes of the Generator: "READY", "ON",
and "DONE". A relationship of the modes and the events required to select each
mode is provided in section 5.1
In the "READY" mode the power level and the count-up timer can be set, the
impedance display is blank, the RF ON/OFF Button is not illuminated, and the
audible tone is off.
In the "ON" mode, which is initiated when the RF ON/OFF Button is pressed, RF
output is active, RF ON/OFF Button/Light is illuminated green, timer display
counts up from 0 seconds to the set time, the audible tone is ON, and the actual
power and impedance are displayed. The RF output is terminated and the mode
changes to DONE (or FAULT) mode when the timer elapses, when the RF
ON/OFF Button is depressed during ON mode, or when an error is detected.
In the DONE mode, the RF output is OFF, the RF ON/OFF Button is not
illuminated, and the audible tone is off. The Generator passes through the DONE
mode as a transition from the ON mode back to the READY mode. The DONE
mode lasts for five seconds and then automatically changes to the READY mode.
The last measured Power, Impedance, and Timer Displays are held during the
DONE mode. The last two seconds of the DONE mode will display an error code
if termination was due to an error.
Note: Two other modes that the Generator has are the POST mode and the FAULT
mode.
The Generator enters into the POST mode when it is first turned on. In the POST mode
the Generator performs a self-test of power generation, measurement, and control
circuitry.
The Generator either transitions to the READY mode upon successful
completion of a self-test or transitions to the FAULT mode if the self-test fails.
The Generator enters into the FAULT mode when a non-recoverable error is detected, at
which point the FAULT light (Fig. 5-1H) is illuminated and a rapid burst of audible tones
is produced.