7.11. TURNING RF POWER "OFF"

RF output is terminated and the mode changes to the DONE mode when either the
timer elapses or when the RF ON/OFF Button is pressed during the ON mode, at
which point the RF ON/OFF Light will turn "OFF".
RF output is also terminated and the mode changes to the FAULT mode when an
error is detected. If the error is recoverable (E001 or E002), the unit proceeds to
the DONE mode.
illumination of the FAULT Indicator and a rapid burst of audible tones, the
Generator must be turned "OFF" and then "ON" again to clear the error, after
which the sequence of initializing steps must be performed again from Section 7.7
Note: If a non-recoverable error is encountered repeatedly, the Generator is not
functioning properly and needs servicing. Contact Baylis Medical Company.

7.12. RE-APPLY RF POWER

To re-apply RF power, repeat steps from section 7.7
For a non-recoverable error, which is illustrated by the
27 DMR RFP-100 3.3 V-13 16-Jan-2012_EN.doc