Generator Software Flow Chart; Figure 5-3 - Software Flow Chart - Baylis Medical RFP-100A Gebrauchsanweisung

Hf-punkturgenerator
Vorschau ausblenden Andere Handbücher für RFP-100A:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5.1.

GENERATOR SOFTWARE FLOW CHART

After 5
seconds in
DONE
mode
Generator
Off
Generator turned on
POST Mode
Self diagnostics
Generator passes
self-tests
READY Mode
SET indicator lit
Time can be
adjusted
Power can be
adjusted
Either
RF ON/OFF button pressed
Footswitch depressed
ON Mode
Energy is delivered
RF ON/OFF
BUTTON/LIGHT lit
TIME display counts up
from 0 seconds
Power display indicates
measured applied power
Impedance display shows
measured Impedance
Self test
Either
DONE Mode (5 seconds)
RF delivery is ceased
TIME displays duration
energy was applied
POWER displays last
measured power delivery

Figure 5-3 - Software Flow Chart

Failed test
Non-recoverable
fault
RF ON button pressed
TIME expired
Footswitch released
Recoverable Fault
15 DMR RFP-100 3.3 V-13 16-Jan-2012_EN.doc
FAIL Mode
Fault indicator lit
Impedance
display flashes
error code
Contact Baylis
Medical Company