Herunterladen Inhalt Inhalt Diese Seite drucken

Error Codes - Baylis Medical RFP-100A Gebrauchsanweisung

Hf-punkturgenerator
Vorschau ausblenden Andere Handbücher für RFP-100A:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

9.4.

ERROR CODES

Summary of Error Conditions and Error Codes
Error Code
H001 (I)
CPU configuration error
H002 (I)
COP time-out error
H003 (I)
ROM CRC error
H004 (I)
RAM error
H005 (I)
Illegal CPU instruction
P001 (I)
RF generation / measurement error
E001 (I)
Low impedance measured
E002 (I)
High impedance measured
E003 (P)
RF power too high
E004 (P)
RF current too high
E005 (P)
RF current measurement error
E006 (P)
RF voltage too high
E007 (P)
RF hardware shutdown
CPU Hardware self-test errors
Reason
POST errors
ON mode errors
34 DMR RFP-100 3.3 V-13 16-Jan-2012_EN.doc
Non-recoverable Recoverable