5.8.

FAULT INDICATOR

The FAULT Indicator is shown in Figure 5-1H. It is a red LED that indicates that
a system error has occurred. The error code is displayed on the IMPEDANCE
Display (Fig. 5-1K). There are three categories of error codes: error(s) during the
POST mode, error(s) during the ON mode, and error(s) in the CPU Hardware
self-test. Recoverable errors do not cause the Fault Indicator to illuminate, but
instead cause a transition to the DONE mode. Non-recoverable errors cause a
transition to the FAULT mode and illuminate the LED. To interpret the error
codes, refer to the table entitled "Summary of Error Conditions and Error Codes"
in Section 9.4. RF power is disabled when the FAULT Indicator is illuminated.
5.9.

POWER SWITCH

The POWER Switch is shown in Figure 5-2A. Turn the Generator "ON" by
pressing the rocker switch to the "I" position. Pressing the rocker switch to the
"0" position will power the Generator "OFF".

5.10. VOLTAGE SELECTOR SWITCH

The orientation of this switch (Figure 5-2D), in conjunction with the Fuse Drawer
(Figure 5-2E), determines the AC input voltage range (100-120V~ or 220-
240V~).
Note: The voltage selector and the orientation of the Fuse Drawer must be set to the
same voltage setting. Do not change the position of the voltage selector while the system
is plugged in.

5.11. FOOTSWITCH

The pneumatic footswitch can be used as an alternative to the RF ON/OFF button
described in section 5.2. As an alternative, it is also responsible for the three
operational modes of the Generator: "READY", "ON", and "DONE". A
relationship of the modes and the events required to select each mode is provided
in section 5.1. Footswitch operation is abbreviated on the label "Generator
Operation" found on the top of the generator and described below.
In the "READY" mode the power level and the count-up timer can be set, the
impedance display is blank, the RF ON/OFF Button is not illuminated, and the
audible tone is off.
In the "ON" mode, which is initiated when the footswitch is pressed and held, RF
output is active, RF ON/OFF Button/Light is illuminated green, timer display
counts up from 0 seconds to the set time, the audible tone is ON, and the actual
power and impedance are displayed. The RF output is terminated and the mode
changes to DONE (or FAULT) mode when the timer elapses, when the footswitch
20 DMR RFP-100 3.3 V-13 16-Jan-2012_EN.doc