SECTION 9: SPECIFICATIONS

9.1.

GENERAL SPECIFICATIONS

Line Power:
Current Rating:
Fuse Rating:
Power Cord Length: 10 feet
Catheter Connector: Quick Connect female 4 pin
Indifferent Electrode Connector: Standard male 2-pin for commercial pads
Recommended Pads: Valley Lab, Pfizer
RS232 Cable Connector: Standard RS232 9-pin female Serial Port Connector
Generator:
Power Output:
Impedance:
Count-up Timer:
100-120 VAC (Domestic Version - CANADA/US)
220-240 VAC (International Version)
1.0A, 60 Hz (100 - 120 VAC) - Domestic Version
0.5A, 50 Hz (220-240 VAC) - International Version
2.0A, 250V, Slow Blow, 14 I²t, UL/ CSA (Domestic
Version).
Recommended 2.0A fuse: Littlefuse, 218 series
(P/N: 0218002.HXP)
1.0A, 250V, Time Lag, 11 I²t, IEC (International Version)
Recommended 1.0A fuse: Littlefuse, 239 series
(P/N: 0239001.MXP)
1-25 Watts, settable in 1-Watt increments
25 Watts into a resistive load of 500-2000 ohms
Maximum output current of 0.224 A RMS
Maximum output voltage of 224.0 V RMS
Into resistive load range of 100-6000 ohms
Measurement accuracy: ± 5 % ± 2 Watts
Measurement range 100-6000 ohms
Measurement accuracy: ± 10 % ± 10 ohms
Settable from 1-99 seconds
Display resolution: 1 second
Accuracy: 0.1 second
29 DMR RFP-100 3.3 V-13 16-Jan-2012_EN.doc