You have selected an outlet port and the flow is lower than 0.8l /min for a period of 10 minutes or more. The product will
display a Check Water Flow Alarm for the given port.
Good flow is reduced to Low flow for > 10 minutes or reduced to No flow for > 15 seconds.
N N o o t t e e
• The alarm displays if the flow is not at an optimal level on each port. This alarm will ask the operator to confirm which
ports are currently in use.
• The addition of a port does not need the operator to confirm the addition.
• The removal of a port requires the operator to confirm.
• The Check Water Flow Alarm for the given outlet port stops, if the operator confirms removal.
• The audio paused timeout is 10 minutes for the check water flow alarm.
N N o o w w a a t t e e r r f f l l o o w w a a l l a a r r m m
This alarm notifies the operator of water flow that is not optimal.
Alarm generation:
When all 3 ports have a flow lower than 0.6l/min for a minimum of 20 seconds, and the pump current is greater than or
equal to 1.0A and the pump is at full speed.
N N o o t t e e
• The pump continues to operate when the No Flow Alarm is active.
• The audio paused timeout is 5 minutes for the now water flow alarm.
N N o o w w a a t t e e r r a a l l a a r r m m
This alarm notifies the operator if there is no water in the system and the pump has shut down.
Alarm generation:
Pump is active in the Manual or Automatic mode. The pump current is lower than 1.0 A and 1 of the ports has a low flow
for more than 60 seconds.
Pump is active in the Manual or Automatic mode. The pump current is lower than 1.0 A and all 3 port flow states are no
flow for more than 60 seconds.
The pump current is less than 0.5 A for 20 seconds.
N N o o t t e e - - The audio paused timeout is 5 minutes for the no water alarm.
P P o o w w e e r r b b a a c c k k u u p p l l e e v v e e l l a a l l a a r r m m
This alarm notifies to the operator an indicator of the status of the Backup Power Level.
N N o o t t e e
• The indicator will remain active until a qualified technician replaces the battery.
• There is no reduction in the usability of the product. The product remains functional and a visual alarm is displays.
• The product will disable the Backup Power Level Alarm in Sleep Mode. Otherwise, the alarm is enabled.
Alarm generation:
The product will display Backup Power Level Alarm when the battery level backup power is less than 100 minutes of
alarms. Once activated, the Backup Power Level Alarm will remain active until you power off the product.
N N o o t t e e - - The audio paused timeout is 10 minutes for the power backup level alarm.