Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker Altrix Bedienungsanleitung Seite 58

Präzisions-temperaturregelsystem
Vorschau ausblenden Andere Handbücher für Altrix:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 201
C C h h e e c c k k p p a a t t i i e e n n t t p p r r o o b b e e a a l l a a r r m m
This alarm notifies the operator that data provided by the probe is not normal or appears removed.
The product only activates the Check Patient Probe Alarm when met during an Active Therapy. Otherwise, the alarm is
disabled.
Alarm generation:
Primary patient temperature changes by more than 1.0° C within two minutes.
N N o o t t e e
• The product will deactivate the heat exchange and keep the pump activated as requested by the Active Therapy.
• The audio paused timeout is 5 minutes for the check patient probe alarm.
P P a a t t i i e e n n t t p p r r o o b b e e m m a a l l f f u u n n c c t t i i o o n n a a l l a a r r m m
This alarm notifies the operator that the probe is not providing information to the product during an active therapy.
Alarm generation:
When the primary patient probe is in a shorted, opened condition, or out of range for more than 30 seconds, the product will
display the Patient Probe Malfunction Alarm.
N N o o t t e e
• The product will deactivate the heat exchange and keep the pump activated as requested by the Active Therapy.
• The audio paused timeout is 5 minutes for the patient probe malfunction alarm.
P P a a t t i i e e n n t t p p r r o o b b e e d d i i s s c c o o n n n n e e c c t t a a l l a a r r m m
This alarm notifies the operator that the probe is not providing information to the product during an active therapy.
Alarm generation:
When the adapter cable for the primary probe is removed and the reading of the Primary Patient probe is out of range for
more than 30 seconds, the product displays the Patient Probe Disconnected alarm.
N N o o t t e e - - The audio paused timeout is 5 minutes for the patient probe disconnected alarm.
P P a a t t i i e e n n t t t t e e m m p p e e r r a a t t u u r r e e d d e e v v i i a a t t i i o o n n m m e e d d i i u u m m a a l l a a r r m m
This alarm notifies the operator that the patient is not responding as expected in the active therapy.
Alarm generation:
The product will display the Patient Temperature Deviation Medium Alarm, if after the initial achievement of the current
patient temperature target, the actual primary patient temperature becomes more than 1.0° C above or below the Current
Target Temperature.
N N o o t t e e - - The audio paused timeout is 30 minutes for the patient temperature deviation medium alarm.
EN
54
8001-109-005 Rev C.2

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis