S S e e l l e e c c t t i i n n g g a a n n d d s s e e t t t t i i n n g g t t h h e e p p r r i i m m a a r r y y p p r r o o b b e e
A patient probe displays when present, stable, and confirmed. The choice of Probe A or Probe B is highlighted when you
insert the cable into port A and port B. If you only insert one cable, the active port displays.
1. Tap the Settings button.
a. In the standby mode, tap the Back button to display the edit settings screen.
b. In an active therapy mode, tap the Next button.
2. Tap Select Probe to display the Select Primary Probe (Probe A or Probe B) screen. If both probes are present, the
default is probe A.
3. Tap A or B, if applicable.
4. Tap Confirm.
N N o o t t e e
• The message "Probe stabilization in progress... Please wait" is displayed.
• When you initially select a probe (A or B), detected is checked. When stabilized, the Ready check is displayed.
• If the probe is not stabilized within three minutes, the message "Probe stabilization error" will appear. Tap the Help
button for more detail.
• You can select help at any time to display help with the current screen or icon descriptions.
F F i i l l l l i i n n g g a a t t h h e e r r m m a a l l t t r r a a n n s s f f e e r r d d e e v v i i c c e e
Read the operations manual for the individual thermal transfer devices for warnings, cautions, and safe operating
instructions before use.
N N o o t t e e - - These instructions are to pre-fill the thermal transfer devices only, not therapy. See Switching modes (page 31).
To fill a thermal transfer device:
1. Connect a thermal transfer device following: Connecting and disconnecting thermal transfer devices (page 23).
2. Lay the thermal transfer device on a flat surface. Make sure the thermal transfer device is flat for water flow.
3. Open all of the clamps on the connector hose and thermal transfer device.
4. Make sure that the controller is powered.
5. Tap the Stand-by button.
6. Tap the Manual mode button.
7. Tap Confirm.
8001-109-005 Rev C.2
A
F F i i g g u u r r e e 1 1 7 7 – – R R e e s s e e r r v v o o i i r r f f i i l l l l l l i i n n e e s s
27
EN