Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker Altrix Bedienungsanleitung Seite 34

Präzisions-temperaturregelsystem
Vorschau ausblenden Andere Handbücher für Altrix:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 201
3. Tap Confirm.
4. Tap the Edit button to make changes.
S S t t a a r r t t i i n n g g M M a a n n u u a a l l m m o o d d e e
In Manual mode, the therapy will cool or warm the water to a selected water target temperature. The operator must observe
the patient's temperature and manually adjust the water temperature to obtain the desired patient temperature.
1. If desired, select and place the sensing end of the patient probe based on hospital protocol. Connect the reusable
adapter cable to port A or port B on the product. See Selecting and connecting a temperature probe (page 21).
2. Prepare the thermal transfer devices to be used for therapy.
3. See Filling a thermal transfer device (page 27).
4. Apply the thermal transfer device to the patient.
5. Tap Manual mode. The default water target temperature is 40.0° C (104.0° F) upon initial entry.
6. Tap Confirm.
7. To select the desired water temperature, tap the Increase or Decrease buttons or hold the button to go faster.
a. To edit water temperature, tap the Edit button.
8. Tap Confirm.
N N o o t t e e
• The controller determines W W a a r r m m i i n n g g or C C o o o o l l i i n n g g therapy based on the selected water target temperature and the
current water temperature.
• In Manual mode, only the water temperature is controlled.
• A temperature probe is not required when operating in Manual mode.
• After the water target temperature is achieved, water temperature is controlled within +/-0.3° C.
S S t t a a r r t t i i n n g g M M o o n n i i t t o o r r m m o o d d e e
In Monitor mode, no therapy will be delivered, only a display of the current patient temperature.
To start Monitor mode:
1. Connect the reusable adapter cable to port A or port B on the controller. Make sure that the probe is fully seated.
2. Apply the sensing end of patient probe to patient based on hospital protocol. Secure the patient probe to reduce the risk
of accidental dislodgment.
3. Tap the Monitor button.
4. Connect the patient temperature probe to the end of the reusable adapter cable. See Selecting and connecting a
temperature probe (page 21).
N N o o t t e e - - If the product senses a patient probe temperature below 36.0° C (96.8° F) or above 38.0° C (104° F), the
normothermia alarm displays and an audible alarm sounds.
5. Tap Confirm. The screen will display the current patient temperature.
EN
30
F F i i g g u u r r e e 1 1 8 8 – – S S e e t t c c u u s s t t o o m m w w a a r r m m i i n n g g r r a a t t e e
8001-109-005 Rev C.2

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis