Foot support self-tapping screws (6) are secure and are not stripped (option)
Upright X-ray cassette holder is in good condition and can be adjusted to fit all X-ray cassettes (P P r r i i m m e e X X option)
No damage to the Fowler skin and foot skin (P P r r i i m m e e X X option)
No damage to the head and foot trays (P P r r i i m m e e X X option)
Bolt and nut through each Fowler pivot are not loose (P P r r i i m m e e X X option)
Product serial number:
Completed by:
Date:
R R e e t t r r a a c c t t a a b b l l e e f f i i f f t t h h w w h h e e e e l l l l u u b b r r i i c c a a t t i i o o n n
T T o o o o l l s s r r e e q q u u i i r r e e d d : :
MPG-3 grease
Bungee cords
P P r r o o c c e e d d u u r r e e : :
1. Raise the product to the highest position.
2. Raise the base hood and support the hood with bungee cords.
3. Apply MPG-3 grease to the spring (A) and roller (B) (Figure 30).
4. Remove the bungee cords and lower the hood.
5. Verify proper operation before you return the product to service.
1105-409-005 Rev E.1
A
B
F F i i g g u u r r e e 3 3 0 0 – – R R e e t t r r a a c c t t a a b b l l e e f f i i f f t t h h w w h h e e e e l l l l u u b b r r i i c c a a t t i i o o n n