To unpack your product, see the unpacking instructions that are attached to the product inside of the shipping crate.
W W A A R R N N I I N N G G
• Always allow the product to reach room temperature before you set up the product or test functional operations.
Permanent product damage may occur.
• Always operate the product when all operators are clear of the mechanisms.
• Always use care when you handle the power cord. Entanglement, damage to the power cord, or potential shock hazards
may occur. If the power cord is damaged, immediately remove the product from service and contact the appropriate
maintenance personnel.
• Always plug the product directly into a grounded, hospital grade wall outlet. You can only achieve grounding reliability
when you use a hospital grade wall outlet. This product is equipped with a hospital grade plug to protect against electric
shock hazard.
Make sure that the product is working before you put the product into service.
1. Apply the brake. Push on the product to make sure that all four casters are locked.
2. Release the brake. Push on the product to make sure that all four casters are unlocked.
3. Raise and lower the litter with the hydraulic lift system.
4. Raise the product to the highest position and put the product in the Trendelenburg position. Make sure that the head end
lowers to the full down position.
5. Raise the product to the highest position and put the product in the Reverse Trendelenburg position. Make sure that the
foot end lowers to the full down position.
6. Apply the fifth wheel and make sure that it guides and pivots the product.
7. Make sure that the siderails raise, lower, and lock in place.
8. Raise and lower the manual Fowler backrest (head end).
9. Raise and lower the manual Gatch (foot end).
For product with the electric lift or electric litter option:
1. Complete the setup checklist above.
2. Plug the product into a grounded, hospital grade wall outlet. Make sure that the LED lights illuminate on the lockout
keypad.
3. Make sure that each function on the patient siderail control option works.
4. Make sure that each function on the foot end nursing control option works.
5. Raise and lower the electric Fowler backrest (head end).
6. Raise and lower the electric Gatch (foot end).
For product with the scale system option-electric litter option (with or without C C h h a a p p e e r r o o n n e e option):
1. Complete the setup checklist above.
2. Charge the batteries.
1105-409-005 Rev E.1
15
EN