Table 5: Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and the
The
ConMed Linvatec
electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the
VP47XX
and
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the
VP47XX
maximum output power of the communications equipment.
Read Maximum
150 kHz to 80 MHz
Output Power
of Transmitter
W
d
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distances
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note 1: E1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2: E2) These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
VP47XX
and
VP47XXF
VP47XXF
Monitors can help prevent electromagnetic interference by
and
VP47XXF
Monitors as recommended below, according to the
Separation Distance According to Frequency of Transmitter
80 MHz to 800 MHz
= [3.5/V1] √
P
V1 = 3V/m
0.12
0.37
1.17
3.70
11.70
E-47
VP47XX
and
VP47XXF
Monitors are intended for use in an
m
800 MHz to 2.5 GHz
= [3.5/E1] √
d
P
E1 = 3V/m
0.12
0.37
1.14
3.70
11.70
Monitors
= [7/E1] √
d
P
E1 = 3V/m
0.23
0.74
2.33
7.37
23.3