Table 4: Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The
ConMed Linvatec
electromagnetic environment specified below. The customer or the user of this
VP47XXF
Monitor should assure that it is used in such an environment.
IEC 60601
Immunity Test
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and/or people.
VP47XX
and
VP47XXF
Compliance Level
Test Level
3 V/m
80 MHz to 2.5 GHz
E-45
Monitors are intended for use in the
Electromagnetic environment
guidance
d
= [3.5 / E1]
80 MHz to 800 MHz
d
P
= [7 / E1]
800 MHz to 2.5 GHz
P
Where
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
should be less than the compliance
level in each frequency range
Interference may occur in the
vicinity of equipment marked with
the following symbol.
E
VP47XX /
P
d
is
a
,
b
.

Werbung