This chapter describes problems that you might encounter while using the Fluent Fluid Management System and how to solve them. For help with
errors not listed in this chapter, please follow the on-screen prompts.
During Setup
• If the In-FloPak is not detected, apply fi rm, even pressure on the In-FloPak to ensure it is seated properly and press 'Next'.
• If the Waste Bag is not detected, ensure tube connection is oriented vertically.
Loss of Suction
• Ensure the hysteroscope outfl ow tubing is connected to the hysteroscope.
• Ensure the buttocks drape tubing is connected to the buttocks drape.
• Ensure the outfl ow tubing is connected to the outfl ow port.
Visibility Issues
• Verify that the fl uid bag clamp is open.
• To provide a tamponade effect, intrauterine pressure may need to be adjusted during the procedure.
• If performing a MyoSure procedure, advance the MyoSure Tissue Removal Device to the fundus and allow circulation of fl uid through the blade
to clear fi eld.
• Ensure the out-fl ow stop cock is fully open.
Poor Uterine Distension
• Avoid over dilation of the cervix. If the cervix is over dilated, use a second tenaculum to seal the cervix.
• Verify the fl uid bag clamp is open.
• Ensure the pressure setting is adequate.
• Ensure the infl ow tubing is not occluded or pinched.
Cutting Issues
• Ensure the MyoSure Tissue Removal Device is fully inserted into the Fluent Fluid Management System.