Herunterladen Inhalt Inhalt Diese Seite drucken

Chapter 13: Electromagnetic Compatibility; Guidance And Manufacturer's Declaration - Hologic Fluent Anwenderhandbuch

Vorschau ausblenden Andere Handbücher für Fluent:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 383
This chapter provides information about the electromagnetic compatibility of the Fluent Fluid Management System.

Guidance and Manufacturer's Declaration

Electromagnetic Emissions
The FLUENT FLUID MANAGEMENT SYSTEM is intended for use in the electromagnetic environment specified below. The customer or the user of
the FLUENT FLUID MANAGEMENT SYSTEM should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Guidance and Manufacturer's Declaration
Electromagnetic Immunity
The FLUENT FLUID MANAGEMENT SYSTEM is intended for use in the electromagnetic environment specified below. The customer or the user of
the FLUENT FLUID MANAGEMENT SYSTEM should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transients / bursts
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
English
Fluent Fluid Management System

Chapter 13: Electromagnetic Compatibility

Compliance
Electromagnetic environment - guidance
Group 1
The FLUENT FLUID MANAGEMENT SYSTEM uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause
interference in nearby electronic equipment.
Group 2
The FLUENT FLUID MANAGEMENT SYSTEM must emit electro-magnetic energy
in order to perform its intended function. Nearby electronic equipment may be
affected.
Class A
The FLUENT FLUID MANAGEMENT SYSTEM is suitable for use in all establish-
ments other than domestic and those directly connected to the public low-volt-
age power supply network that supplies buildings used for domestic purposes.
Complies
NOTE: The EMISSIONS characteristics of this equipment make it suitable
for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a
Complies
residential environment (for which CISPR 11 class B is normally required) this
equipment might not offer adequate protection to radio-frequency commu-
nication services. The user might need to take mitigation measures, such as
relocating or re-orienting the equipment.
IEC 60601
Compliance level
test level
± 8 kV contact
± 8 kV contact
± 15 kV air
± 15 kV air
± 2 kV for power supply
± 2 kV for power supply
lines
lines
± 1 kV for input / output
lines
± 1 kV line(s) to line(s)
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 2 kV line(s) to earth
0% UT (100% dip in the
0% UT (100% dip in the
UT) for ½ cycle at 0°,
UT) for ½ cycle at 0°,
45°, 90°, 135°, 180°,
45°, 90°, 135°, 180°,
225°, 270°, and 315°
225°, 270°, and 315°
0% UT (100% dip in the
0% UT (100% dip in the
UT) for 1 cycle and 70%
UT) for 1 cycle and 70%
(30% dip in the UT) UT
(30% dip in the UT) UT
for 25/30 cycles at 0°
for 25/30 cycles at 0°
0% UT (100% dip in the
0% UT (100% dip in the
UT) for 250/300 cycles
UT) for 250/300 cycles
30 A/m
30 A/m
Chapter 13: Electromagnetic Compatibility
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
If the user/operator of the FLUENT FLUID
MANAGEMENT SYSTEM requires continued
operation during power mains interruptions,
it is recommended that the FLUENT FLUID
MANAGEMENT SYSTEM be powered from an
uninterruptible power supply or battery.
Power frequency magnetic fields should be at lev-
els characteristic of a typical location in a typical
commercial or hospital environment.
45

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis