Parameter Level 2
Values in Parameter Level 2 can only be adjusted after the password "9403" was entered. The password cannot be changed.
Entering password: Press "Enter", press "Select
Handpiece activation
Handpiece existing/HP 02
Handpiece existing/HP 03
Handpiece existing/HP 04
Handpiece existing/HP 05
Handpiece
Max. Speed
Handpiece max speed/HP 01
Handpiece max speed/HP 02
Handpiece max speed/HP 03
Handpiece max speed/HP 04
Handpiece max speed/HP 05
AS-Zone
for handpieces
Handpiece AS-Mode/HP 06
Handpiece AS-Mode/HP 07
Handpiece AS-Mode/HP 08
Handpiece AS-Mode/HP 09
Pump parameters
Pump/Backwards
turn mode variable
Pump/Way backwards
Pump/Speed backwards
Pump/Range 1 increment
Pump/Range 1 end
Pump/Range 2 increment
Pump/Range 2 end
Pump/Range 3 increment
Pump/Range 3 end
MD 11 Operating Instructions, Art. No. 31665, V10/14
▲ or ▼", (
for fast forward or backward keep key pressed).
Name of handpiece
Choice
on display
16 : 1
yes/no
20 : 1
yes/no
32 : 1
yes/no
70 : 1
yes/no
Name of handpiece
Speed range
on the display
rpm
1 : 1
300 – 40,000
16 : 1
20 – 2400
20 : 1
15 – 1700
32 : 1
10 – 1000
70 : 1
5 – 600
Handpiece
Speedrange
on display
... to ... rpm
16 : 1
20 – 45
20 : 1
15 – 45
32 : 1
10 – 45
70 : 1
5 – 45
Range
Factory
default
No/Yes
Yes
1 – 100%
25%
10 – 50%
33%
1 – 10%
5%
5 – 50%
10%
1 – 10%
5%
10 – 90%
50%
1 – 10%
10%
20 – 100%
100%
Factory
Definition
default
no
Only one handpiece or contra angle can
be selected.
yes
no
no
Factory
Definition
default
40,000
Limit the maximum speed of your
handpieces according to your own
2400
experience.
1700
1000
600
Factory
Definition
default
20
Effective AS-range 20 – 45 rpm
20
Effective AS-range 15 – 45 rpm
20
Effective AS-range 10 – 45 rpm
20
Effective AS-range 5 – 45 rpm
Definition
The pressures in the tube set vary according to the
pump speed. In "variable mode" this is considered, to
prevent of spilling when pump is switched off in
backwards mode.
Specify how far the pump turns backwards
Specify how fast the pump has to turn backwards to
prevent of spilling after the switch off of the motor.
Adjustment steps in section 1.
Set the range where section 1 is active
Adjustment steps in section 2
Set the range where section 2 is active.
Adjustment steps in section 3
Set the range where section 3 is active.