Herunterladen Inhalt Inhalt Diese Seite drucken

After Starting The First Centrifugation Run, Enter The Maximum Permitted Number Of Cycles Or Deactivate The Cycle Counter; Reset Cycle Counter To "0" And Enter The Maximum Permitted Number Of Cycles - Hettich ROTINA 380 Robotic Bedienungs- Und Einbauhinweise

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

24.1
After starting the first centrifugation run, enter the maximum permitted number of cycles or deactivate
the cycle counter

Enter max cycles = 30000 is displayed.
Use the knob
to set the maximum permitted number of rotor cycles specified on the hanger.
For rotors and hangers which are not labelled with the maximum permissible number of running cycles, the cycle
counter can be deactivated. Turn the knob
deactivated).

Press the
key to save the setting.
START
Briefly, Store max cycles ... is displayed as confirmation.
24.2

Reset cycle counter to "0" and enter the maximum permitted number of cycles

If the rotor is at standstill, this can be set as follows:
You can scroll backwards through the menu by pressing the
The operation can be cancelled at any time by pressing the
stored.

Keep the
key held down for 8 seconds.
PROG
After 8 seconds,  Machine Menu  appears in the display.

Press the
key as often as necessary until the ->Operating Time is displayed.
PROG

Press the
key. The external operating hours are displayed, e.g. OP Time ext = 0h25m.
START

Press the
key as often as is required until the rotor cycles are displayed, e.g. Cycles = 30001 of 30000.
PROG

key. The number of rotor cycles is displayed in parentheses áñ , e.g. Cycles = 30001 of
Press the
RCF
30000.

Turn the knob
to the left to reset the number of rotor cycles to "0".
If the rotor cycles are not reset to "0", press the
to be displayed and the setting is not saved.

key. The maximum permitted number of rotor cycles is displayed in parentheses , e.g. Cycles
Press the
RCF
= 0 of 30000.

Use the knob
to set the maximum permitted number of rotor cycles specified on the hanger.

Press the
key to save the settings.
START
In confirmation, Store cycles ... is displayed briefly. Next the rotor cycles are displayed, e.g. Cycles = 0 of
30000.

Press the
key twice to exit the "Operating Time" menu or press the
OPEN / STOP
exit the "Machine Menu".
to the left until disabled is displayed (disabled = cycle counter is
key.
T/°C
OPEN / STOP
key, which causes Max cycles = actual cycles
START
key. In this case, the settings are not