EN
30
Centrifuge address
Ex works, the address is ] = 29. Address is set.
31
Query the operating hours, centrifugation runs and cycle counter
The operating hours are divided up into internal and external operating hours.
Internal operating hours: Total time the device was switched on.
External operating hours: Total time of previous centrifugation runs.
The query is performed as follows when the rotor is at a standstill:
You can scroll backwards through the menu by pressing the
Keep the
key held down for 8 seconds.
PROG
After 8 seconds, Machine Menu appears in the display.
Press the
key as often as necessary until the -> Operating Time is displayed.
PROG
Press the
key. The external operating hours are displayed, e.g. OP Time ext = 0h25m.
START
Press the
key. The internal operating hours are displayed, e.g. OP Time int = 1h36m.
PROG
Press the
key. The number of all centrifugation runs is displayed, e.g. Number of Starts = 10.
PROG
Press the
key. The number of rotor cycles (centrifugation runs) of the current rotor code since the last time
PROG
the cycle counter was reset to "0" and the permitted number of rotor cycles are displayed, e.g. CYCLES = 5120
of 30000.
Press the
key. The number of all rotor cycles (centrifugation runs) of the current rotor code is displayed,
PROG
e.g. Rotor cycles total = 37490. This value cannot be set.
Press the
key twice to leave the menu, "Operating Time" or press the key
OPEN / STOP
to leave the "Machine Menu".
32
Query the system information
The following system information can be queried:
Centrifuge model
Mains voltage
Rotor information
Program version of the centrifuge
Program version of the frequency converter
The system information can be queried as follows when the rotor is at a standstill:
You can scroll backwards through the menu by pressing the
Keep the
key held down for 8 seconds.
PROG
After 8 seconds, Machine Menu appears in the display.
Press the
key as often as necessary until the ->Info is displayed.
PROG
Press the
key. The centrifuge model is displayed.
START
Press the
key. The mains voltage is displayed, e.g. Mains Voltage : 230 V.
PROG
Press the
key. The rotorcode (rotor), the maximum rotor speed (Nmax) and the preset centrifuging radius
PROG
(R) of the last rotor that was recognized by the rotor recognition , e.g. Rotor 4: Nmax = 4500 R=184.
The last recognized rotor is marked with a star (). Use the adjusting knob
permitted in the centrifuge displayed.
Press the
key. The program version of the centrifuge is displayed, e.g. SW version = 01.00.
PROG
Press the
key. The program version of the frequency converter is displayed e.g. FC-SW-Version = 4.
PROG
Press the
key twice to leave the "Info" menu or press the key
OPEN / STOP
"Machine Menu".
33
Immediate display of centrifugation data after switch-on.
Switch on the power supply. (Switch setting ).
The first time the display changes optically, (inverse display) press any key and keep it held down. The
centrifugation data is displayed immediately.
78/203
key.
T/°C
OPEN / STOP
key.
T/°C
to have the information of the rotors
three times to leave the
OPEN / STOP
three times