Calling up program linkage


Press the
button repeatedly until the parameter RCL A...Z is displayed.
PROG

Use the adjusting knob
Press the
key. Briefly, Multi program recall ... is displayed as confirmation.
START

The centrifugation data of the first program of the program linkage as well as the total runtime of the program
linkage are displayed.
19.5

Automatic temporary storage

The program place 0 serves as temporary storage for centrifugation data of the last centrifugation run that took place.
No programs can be saved at this program place.
After every start to a centrifugation run, the centrifugation data that is used for the run is automatically saved at
program place "0" and can be accessed there.
20

Centrifugation

The centrifuge may only be operated in an automated system with loading and unloading by a robot.
The manufacturer of the automated system is responsible for its safety.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous substances or
objects may be within the safety margin of 300 mm around the centrifuge.
The centrifuge can be operated via the operating unit (key position "0") or controlled via the interface (key
position "LOCK 2"). The functions of the various key positions are described in the "Key-operated switch"
chapter.
A centrifugation run can be stopped at any time by pushing the key
During a centrifugation run, parameters can be selected and modified (see chapter, "Change settings during
centrifugation run").
Via the interface or with the buttons
time. Switching with the buttons
display is worked with, the centrifuging radius must be input.
If Enter max cycles = 30000 is displayed, first of all the maximum permitted number of rotor cycles specified
on the hanger is entered, before the centrifugation run can be restarted (see chapter "Cycle Counter").
If the rotor has been changed, no centrifugation run takes place and this display is shown, e.g.
Rotor 4 Nmax= 4500 R=184 mm (see chapter "Rotor Identification").
Operation errors and malfunctions will be shown (see Chapter "Faults").
20.1

Centrifugation with preselected time


Set the centrifugation parameters or call up a program or a program linkage (see the chapters on "Enter
centrifugation parameters", "Calling up programs" or "Program linkage").

Start the centrifugation run via the interface. The LED in the
it is subsequently lit.

After the time elapses or if the centrifugation run is aborted via the interface or by pressing the
the run-down occurs with the activated run-down parameters. The run-down parameter is displayed e.g.
The right-hand LED in the
key and the right LED in the
During the centrifugation run, the rotor speed or the RCF value, the temperature in the centrifuging chamber (only for
centrifuge with cooling), and the remaining time are displayed.
66/83
to set the program place you want.
and
RPM
RCF
and
is not possible if program links are worked with. If the RCF
RCF
RPM
key is lit. Once the rotor has come to a standstill, the LED in the
OPEN / STOP
key switch off and the left LED in the
OPEN / STOP
OPEN / STOP
, one can switch between the RPM and RCF display at any
key blinks until the rotor has been imported,
START
OPEN / STOP
.
key,
OPEN / STOP
9.
START
key lights up.