RS 232/RS 485 Communication Description

Both the RS 232 and the RS 485 interfaces are available on the
connector J2
J2 – Serial Port
Tab. 6
PIN N.
1
2
3
4
5
6
7
8
9
The communication protocol is the same (see the structure below),
but only the RS 485 manages the address field. Therefore to enable
the RS 485 is necessary to select the type of communication as well
as the device address by means of the T-plus software.
The RS-485 is a 2-wire (gnd optional) half-duplex communication link.
NOTE

Communication Format

MS40+ Single Stage Rotary Vane Pumps User Manual / 87-901-008-01
SIGNAL NAME
+ 5 V out
TX (RS232)
RX (RS232)
Spare
GND
A+ (RS485)
Spare
B- (RS485)
Spare
8 data bit
no parity
1 stop bit
baud rate: 600/1200/2400/4800/9600 programmable