Single stage rotary vane pumps

Werbung

5
Technical Information

Maintenance Actions

Maintenance Actions

Maintenance may be seen as the totality of all scheduled and
unscheduled maintenance work.
Tab. 2
Oil level checking
Oil change
Exhaust filter replacement
Anti such-back valve checking
Fan cover cleaning
The scheduled maintenance is supported by an automatic timer provided by the
NOTE
electronic controller. After 8000 h of operation the status led starts blink-ing
(400 ms period) orange. The led blinking highlights that is necessary to change
the oil. To switch off the alarm it's necessary to use the serial interface (see
window list in the following pages). It is possible to read the Scheduled
Maintenance timer by using the serial interface.
The frequency with which repairs are performed depends on the process and
NOTE
presence of substances that shorten pump life (dust, abrasives, solvents, water,
chemically aggressive substances).
The pump must be cleaned at regular intervals of time.
Do not clean with Alcohol the plastic or rubber components of the pump.
CAUTION!
Use only the strictly necessary amount of lubricant; an excess of
lubricating oil, like when there is none, may sometimes compromise
proper operation of the pump.
64/90
MS40+ Single Stage Rotary Vane Pumps User Manual / 87-901-008-01
SCHEDULED MAINTENANCE: Maintaining the nominal state of
operation.
UNSCHEDULED MAINTENANCE: Restoring the nominal state
of operation.
Daily (before every starting)
8.000 hours (light applications)
If oil mist at exhaust or yearly
6 months
6 months
(E)