Electromagnetic Compatibility

Like other electrical medical equipment, the
to ensure electromagnetic compatibility with other electrical medical devices. To ensure
electromagnetic compatibility (EMC), the AMM215WTD must be installed and operated
according to the EMC information provided in this manual.
Note
The AMM215WTD has been designed and tested to comply with IEC 60601-
1-2:2001 requirements for EMC with other devices.
Caution Portable and mobile RF communications equipment may affect
the normal function of the AMM215WTD.
Warning Do not use cables or accessories other than those provided
with the AMM215WTD, as this may result in increased electromagnetic emis-
sions or decreased immunity to such emissions.
Warning
If the AMM215WTD is used adjacent to or stacked with other
equipment, observe and verify normal operation of the AMM215WTD in the
configuration in which it will be used prior to using it in a surgical procedure.
Consult the tables below for guidance in placing the AMM215WTD.
Manufacturer's declaration - electromagnetic emission
The Model AMM215WTD is intended for use in the electromagnetic environment specified below.
The customer or the user of AMM215WTD should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage Fluctuations
IEC61000-3-3
Manufacturer's Declaration - Electromagnetic Immunity
The Model AMM215WTD is intended for use in the electromagnetic environment specified below.
The customer or the user of AMM215WTD should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC61000-4-2
Electrical fast transient
/burst IEC61000-4-4
Surge
IEC61000-4-5
Compliance
The Model AMM215WTD uses RF energy only for its internal
Group 1
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment
Class B
AMM215WTD is suitable for use in all establishments, including
Class D
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes
Complies
IEC 60601 Test Level
6 kV contact
8 kV air
2 kV for power supply lines
1 kV for input / output lines
1 kV differential mode
2 kV common mode
AMM215WTD
requires special precautions
Electromagnetic environment - guidance
Compliance Level
6 kV contact
8 kV air
2 kV for power supply lines
1 kV for input / output lines
1 kV differential mode
2 kV common mode
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.