Electrical connection of the P2VA1/A2
16
Notice
The 0 V power supply (pin 3) and 0 V output (pin 2) lines
are linked internally.
Important
The 1-wire TEDS Technology can not readout from the
QuantumX system.
The transducer is designed for direct voltage operation
(15 to 30 volts).
The circuit is intended for operating with a separated
extra-low voltage (SELV circuit). It is not intended for
connection to a dc network in accordance with
EN 61010-1.
Should the equipment be operated on a dc voltage net­
work, additional precautions must be taken to discharge
excess voltages.
Notice
The P2V meets EMC guideline requirements (a condition
of the transducer being granted the CE mark).
To prevent measured value errors when there is heavy
high-frequency interference, a shielded connection cable
with a metal female cable connector must be used, with
the cable shielding connected to the entire surface of the
female cable connector. The enclosed HBM connection
cable meets this requirement.