Herunterladen Inhalt Inhalt Diese Seite drucken

Scope Of Delivery; Instructions For Use; Preparation For Use - Ottobock 1M10 Adjust Gebrauchsanweisung

Prothesenfuß
Vorschau ausblenden Andere Handbücher für 1M10 Adjust:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1

4 Scope of delivery

Quantity
1
1
1
Spare parts/accessories (not included in the scope of delivery)
Fig.
1
Footshell
1
Insert for function module

5 Preparation for use

CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
Observe the alignment and assembly instructions.
INFORMATION
Not all of the materials may be available in your country. In this case, please contact the loc­
al branch of the manufacturer to obtain information on alternative materials.
5.1 Alignment
5.1.1 Applying/removing the footshell
Inserting the foot into the footshell
1) Slide the prosthetic foot into the footshell (see fig. 3, see fig. 4).
2) Press the heel of the prosthetic foot into the footshell until it engages (see fig. 5).
Removing the footshell
1) Push the lock of the footshell to the rear and pull up the heel of the prosthetic foot.
2) Remove the prosthetic foot from the footshell.
5.1.2 Bench Alignment
TT bench alignment
Required materials: 662M4 goniometer, 743S12 heel height measuring apparatus, 743A80 50:50 gauge,
alignment apparatus (e.g. 743L200 L.A.S.A.R. Assembly or 743A200 PROS.A. Assembly)
Perform the assembly and alignment of the prosthetic components in the alignment apparatus according to
the following specifications:
Sagittal plane
Heel height: Effective heel height (shoe heel height - sole thickness of forefoot) + 5 mm
   
Exterior foot rotation: approx. 5°
   
Anterior placement, middle of the prosthetic foot to the alignment reference line: 30 mm
   
   
Connect the prosthetic foot and prosthetic socket using the chosen adapters. The instructions for
use of the adapters must be observed.
   
Determine the centre of the prosthetic socket with the 50:50 gauge. Align the prosthetic socket
centrally to the alignment reference line.
Socket flexion: Individual residual limb flexion + 5°
Frontal plane
12
Designation

Instructions for use

Prosthetic foot
Protective sock
Reference number
Bench alignment process
Reference number
SL=Spectra-Sock-7
Reference number
2C1
2D11

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis