Herunterladen Inhalt Inhalt Diese Seite drucken

DeVilbiss SmartLink DV5M Bedienungsanleitung Seite 12

Modul
Vorschau ausblenden Andere Handbücher für SmartLink DV5M:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 38
Immunity Test
Power Frequency
50/60Hz Magnetic
Field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
For transmitters rated at a maximum output power not listed above, the recommended separation distance
D in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
EN - 12
IEC 60601 Test
Level
3A/m
3 Vrms from
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance Level
3A/m
V1 = 3 Vrms
E1 = 3V/m
Electromagnetic Environment -
Guidance
Power frequency magnetic fields
should be that of a typical location in
a typical commercial or hospital
environment.
Portable and mobile RF
communications equipment should be
separated from the device by no less
than the recommended separation
distances calculated/listed below:
D=(3.5/V1)√ P
D=(3.5/E1)√ P
80 to 800 MHz
D=(7/E1)√ P
800 MHz to 2.5 GHz
Where P is the maximum power
rating in watts and D is the
recommended separation distance in
meters.
Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey, should
be less than the compliance levels
(V1 and E1).
Interference may occur in the vicinity
of equipment containing a transmitter.
A-DV5M-1

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis