If you need assistance with a technical problem, make sure to fill the complete
checklist (all 23 points!) and fax a copy to Consort: +32/14/429179.
1. purchase date:
2. indicate model:
C831 - C832 - C833 - C835 - C838
5. organisation:
6. name:
7. division:
8. address:
9. postal code:
11. telephone:
13. e-mail:
14. type of connected electrode:
15. store all conductivity cells in distilled water with some detergent overnight.
16. reset the meter (switch on while holding MODE pressed).
17. leave the EC input open, select the conductivity mode and read display.
18. connect a 1000 ohm resistance to the conductivity input and read display.
19. select the °C mode, measure temperature of solutions and read the display.
20. rinse the conductivity cell with a 0.01 M KCl calibration solution (1413 µS).
21. connect the conductivity cell, dip in fresh cal. solution (stir!) and read display.
22. rinse a new conductivity cell with a 0.01 M KCl calibration solution (1413 µS).
23. connect this conductivity cell, dip in fresh cal. solution (stir!) and read display.
C831 • C832 • C833 • C835 • C838
2. purchased from:
10. city:
12. telefax:
Perform the following steps:
Technical assistance form - Conductivity
4. serial no:
yes?
yes?
µS:
µS:
°C:
yes?
µS:
yes?
µS:
71 / 74