If you need assistance with a technical problem, make sure to fill the complete
checklist (all 23 points!) and fax a copy to Consort: +32/14/429179.
1. purchase date:
2. indicate model:
C831 - C832 - C833 - C835 - C838
5. organisation:
6. name:
7. division:
8. address:
9. postal code:
11. telephone:
13. e-mail:
14. type of connected electrode:
15. store all pH electrodes in a 3...4 KCl solution overnight.
16. reset the meter (switch on while holding MODE pressed).
17. short-circuit the pH input, select the mV mode and read display.
18. select the °C mode, measure temperature of solutions and read the display.
19. select the mV mode (do not calibrate!).
20. connect the pH electrode, rinse, dip in fresh buffer 7 (stir!) and read display.
21. rinse the pH electrode, dip in fresh buffer 4 (stir!) and read display.
22. connect a new pH electrode, rinse, dip in fresh buffer 7 (stir!) and read display.
23. rinse this pH electrode, dip in fresh buffer 4 (stir!) and read display.
C831 • C832 • C833 • C835 • C838
2. purchased from:
10. city:
12. telefax:
Perform the following steps:

Technical assistance form - pH

4. serial no:
yes?
yes?
mV:
°C:
yes?
mV:
mV:
mV:
mV: