1. Select the desired mode (S/cm) by pressing é ê . The display will immediately
show the measured value according to the previous calibration. Should you
want to recalibrate, press CAL.
2. The instrument permits to choose between the standards in memory (1413
µS/cm, 12.88 mS/cm, 111.8 mS/cm + max. 3 user tables). Select the proper
values and press OK. The unused standards should be switched off.
3. After rinsing the electrode several times with the first standard solution,
immerse it in that standard. The solution temperature is not so critical but
should lie between 0°C and 30°C. When no Pt1000 is used, do not forget
to compensate manually first! Select [CALIBRATE], press OK and follow the
instructions on the screen until the calibration is finished.
4. Rinse the electrode several times with the sample, immerse it in that solution
and read the display.
5. Rinse the electrode always after use and store it in distilled water (add some
detergent to keep the spongy platinum surface in perfect condition).
When starting a titration, hold MODE pressed and press ê to lock the actual
range and avoid cross-over errors due to differences in the measuring fre-
quency. Hold MODE pressed and press ê to return again to the normal meas-
uring mode.
Temperature compensation:
1. Press CAL.
2. Select [REF.] and press OK.
3. Select the temperature to which all future measurements will be referred to
(25°C or 20°C) (C305x: 20°C...40°C in 1°C steps) and press OK.
Capacitive compensation:
1. Capacitive compensation increases accuracy in the very low conductivity rang-
es (<10 µS/cm). Select whether or not this compensation should be applied.
Verify if the attached electrode is completely dry and press CAL.
2. Select [CAP.COMP] and press OK.
3. Select [YES] or [NO] and press OK. Follow the instructions on the screen.
Resistivity measurement:
1. Select the desired mode (Ω.cm) by pressing é ê . The display will immediately
show the measured value.
Salinity measurement:
1. Select the desired mode (SAL) by pressing é ê . The display will immediately
show the measured value.
TDS measurement:
1. Select the desired mode (mg/l)(TDS) by pressing é ê . The display will imme-
diately show the measured value.
To browse between conductivity, resistivity, salinity or TDS, hold MODE
pressed and press é sequencially.
1. After calibrating and rinsing the ORP and conductivity electrodes with distilled
water, immerse them in the solution to be measured.
2. Select the multi-channel display by pressing é ê and read the display.
C301x • C302x • C303x • C304x • C305x • C306x

Conductivity measurement

µW measurement
(C305x)