CAUTION
Independent manipulation of system components
This may lead to falling caused by:
Malfunctioning of the orthosis.
Breakage of load-bearing components.
→ The orthosis joint may only be opened and repaired by authorised Ottobock technicians.
→ For safety reasons, battery service may only be performed by Ottobock Service (batteries
may not be replaced by the user).

4.2.2 Notes on Transportation

CAUTION
Mechanical influences and/or strain during transportation
Falling or skin irritation may result, caused by:
Malfunctioning of the orthosis.
Breakage of load-bearing components.
Damage to the battery.
Hydraulic unit defects with associated leakage of liquid.
→ Always use the transport packaging for transportation.

4.2.3 Notes on the Power Supply/Battery Charging

CAUTION
Incorrect battery charging
This may lead to falling caused by:
Altered damping behaviour of the orthosis.
Malfunctioning of the orthosis.
→ Observe the charging process instructions (see Page 42).
→ To charge the battery, only use battery chargers that have been approved by Ottobock.
CAUTION
Self-discharge of the battery when the orthosis is not used for extended periods of time
This may lead to falling caused by:
Altered damping behaviour after switching to empty battery mode, as a result of the supply
of electricity to the orthosis electronics is no longer sufficient.
When the orthosis is operating with insufficient battery charge level, it may switch to empty
battery mode. The damping behaviour (high/low) can be chosen by the CPO when program­
ming the settings.
→ We recommend recharging prior to each use (see Page 43).
CAUTION
Charging the orthosis without taking it off
This may lead to falling caused by:
Altered damping behaviour of the orthosis. Swing phase will not initiate because the orthosis
does not function during charging.
Walking and getting caught on the connected battery charger.
→ For safety reasons, remove the orthosis prior to charging it.