7. Commissioning, operation
In order to switch the instrument into "SWITCH-TEST" mode, follow the instructions on
the previous page.
GB
The following is a more detailed explanation of point "2. Preparation for SWITCH-
TEST"
a) Voltage supply for test item (on/off) [If no external supply is required for the
test item, "OFF" should be selected to conserve energy]
b) Unit and resolution (submenu)
XXX
◁▷ ◀▶
0.00
ENTER
52
The switch-test function is not suitable for electronic switches (e.g.
PNP- or NPN switches), but is only for mechanical, potential-free
switches.
Select and confirm (standard units) via
Customer-specific unit; with respect to bar
(input via numeric keypad)
Display resolution in operating mode
(back with
The calculated measured values of the two switching points and the
hysteresis can be reset by pressing the "0" key.
Current cursor position; alter via
Parameter selection from list or menu via
Parameter input via numeric keypad
Input confirmation
WIKA operating instructions ProcessCalibrator, model CPH6000
)
Clear entry
CLEAR