Herunterladen Inhalt Inhalt Diese Seite drucken

During Charging - WIKA CPH6000 Betriebsanleitung

Der prozesskalibrator typ cph6000 vereint die vorteile eines kompakten hand-held- instrumentes mit der umfangreichen funktionalität eines labor-kalibrierinstrumentes. hierdurch gestalten sich alltägliche aufgaben im feld wie messen, prüfen oder kalibrie
Vorschau ausblenden Andere Handbücher für CPH6000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4. Design and function
GB
When the battery charger is no longer being used, the mains plug should be
disconnected from the mains socket. Do not leave the battery charger connected to
the rechargeable battery for longer than one day, since overloading can shorten its
service life.
Should the rechargeable battery still not be fully charged after 24 hours, contact the
manufacturer. When not being used, a fully-charged battery will lose its charge over
time.
Extreme temperatures have an adverse effect on battery charging. As a result, the
battery may first need to be either cooled or warmed, as appropriate.
When the battery is nearly empty, the message "low BAT" appears in the display.
With 0 % battery level, the equipment automatically switches itself off and must then
be recharged using the battery charger.

4.5.1 During charging

4.5.2 Charging the Lithium-Ion batteries
24
When the mains lead/battery charger is connected to the CPH6000,
the battery will be charged, even if the CPH6000 is switched off.
CAUTION!
The battery level during storage or shipping should be between 25 and
50 %.
CAUTION!
The temperature range over which the Lithium-Ion battery can
be charged is 10 ... 45 °C. Charging the Lithium-Ion battery at
temperatures outside of this range may lead to heating or damage. In
addition, the performance of the Lithium-Ion battery can be affected
and the service life reduced.
WARNING!
In order to charge the Lithium-Ion battery, never use any device other
than that specified by WIKA. When the Lithium-Ion battery is used in
devices other than the specified device, the performance and service
life of the Lithium-Ion battery may be reduced, and, should the device
cause an abnormal current to flow, it can cause the Lithium-Ion battery
to become hot, explode or ignite and result in serious injury.
WIKA operating instructions ProcessCalibrator, model CPH6000

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis