7. Commissioning, operation
d)
Functions: Operating functions such as
Tare: Offset correction of the reference pressure value
Min/Max: Minimum/Maximum memory
Alarm: Min/Max alarm (visual and audible)
Filter: Damping/smoothing of the reference sensor signal
For further information see chapter 7.4.1 "SETUP additional menu items: Functions"
e)
CPH-Info: General CPH6000 instrument data such as
Calibration dates for the electrical measurement inputs
Firmware number
Serial No. of the device
For further information see chapter 7.4.2 "SETUP additional menu items: CPH-INFO"
f)
ReferenceSensor: Data for the currently-connected reference pressure sensor such
as
Measuring range
Accuracy class
Sensor measurement type
Information in the event of reference sensor overpressure
Calibration dates for the reference sensor
For further information see chapter 7.4.3 "SETUP additional menu item: Reference-
Sensor"
g)
Ref. Sensor-list:
List of the stored reference sensors that can be attached and are calibrated.
For further information see chapter 7.4.4 "SETUP additional menu item: Ref. Sensor-
list"
h)
CPH-Configuration:
Info: on battery level
Setting options from: Menu language, system time / system clock, display brightness,
powersave function (automatic energy saving mode; see chapter 7.4.5 "SETUP addition-
al menu item: CPH-Configuration")
Input options:
- the ambient temperature during the calibration
- the height difference between reference pressure sensor and test item
(see chapter 6.1 "Requirements for test assemblies with the CPH6000").
For further information see chapter 7.4.5 "SETUP additional menu item: CPH-Config-
uration"
i)
Interface
Switching between USB and RS-232 interfaces incl. entry of baud rate
For further information see chapter 7.4.6 "SETUP additional menu item: Interface"
j)
CLEAR CalData:
Delete all stored calibration data (clear and reset all memory locations)
For further information see chapter 7.4.7 "SETUP additional menu items: CLEAR
CalData"
k)
CPH switch off:
Switching off the CPH6000 ProcessCalibrator
For further information see chapter 7.2.2 "Switching off the model CPH6000
ProcessCalibrator"