Herunterladen Inhalt Inhalt Diese Seite drucken

Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Storz Telecam SL II Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

70
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
Guidance and manufacturer's declaration – electromagnetic immunity
®
TELECAM
SL II model 20 2130 20 is intended for use in the electromagnetic environment specified below. The customer or user of the TELECAM
model 20 2130 20 should ensure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
IEC 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
Voltage dips, short
interruptions and voltage
<5% U
variations on power supply
(>95% dip in U
input lines
for 0.5 cycle
IEC 61000-4-11
40% U
(60% dip in U
for 5 cycles
70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
Appendix
Electromagnetic Compatibility (EMC)
Information
Table 202
Compliance level
Complies
± 6 kV contact
± 8 kV air
Complies
± 2 kV for power supply
lines
± 1 kV for input/output lines
Complies
± 1 kV differential mode
± 2 kV common mode
Complies
*
<5% U
*
T
T
)
(>95% dip in U
T
T
for 0.5 cycle
Complies
40% U
T
T
)
(60% dip in U
)
T
T
for 5 cycles
Complies
70% U
T
T
)
(30% dip in U
)
T
T
for 25 cycles
Complies
<5% U
T
T
)
(>95% dip in U
T
T
for 5 sec
Complies
3 A/m
Anexo
Indicaciones sobre compatibilidad
electromagnética (CEM)
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment. If image
distortion occurs it may be necessary to further filter
the power quality by using a AC line filter.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
)
If the user of the equipment or system requires
continued operation during power mains interruptions,
it is recommended that the equipment or system be
powered from an uninterruptible power supply or a
battery. If image distortion occurs it may be necessary
to further filter the AC input to the TELECAM
model 20 2130 20 by using an external AC line filter.
)
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
®
SL II
®
SL II

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis