10. Expressing your breast milk

CAUTION
To avoid health risks and reduce the risk of injury:
" Milk expression should not be uncomfortable. Very powerful vacuum, misaligned nipple or
incorrect breast shield size can cause trauma to the nipple/breast and decrease milk output.
Adjust the breast shield size (see chapter 7) and vacuum as needed.
14
" Stop using the breast pump if discomfort persists or in the event of negligible milk output. Please
contact your healthcare professional or breastfeeding specialist.
" If you experience pain during pumping, break the seal between the breast and the breast shield
with your finger and remove the breast pump from your breast. 
Information
" Stimulation phase is a fast pumping mode to stimulate milk flow.
• Expression phase is a slower pumping mode for gentle and efficient milk removal.
• Maximum Comfort Vacuum (MCV) is the highest vacuum setting where pumping still feels
comfortable. This is different for every mother and will result in the most efficient milk expression.
" When in Expression phase, increase the vacuum with the + button until pumping feels slightly
uncomfortable (not painful). Then decrease the vacuum with one press of the 3 button. This is
your MCV level.
" It is important to identify the MCV during the first pumping session as this vacuum level will be
remembered by the breast pump. The breast pump will adjust the vacuum automatically to the
MCV level during the next session. While the breast pump adjusts to the MCV level, the +
button will be flashing.
" Consider reassessing your MCV each week to ensure you are pumping at your optimal
vacuum level.

10.1 Operating your breast pump

The Magic InBra
#
are wearing supports the breast pump securely in place while avoiding excessive compression on
your breast. Try loosening the bra strap or switching to a more stretchy bra if you experience
discomfort.
1.
breast pump may be worn with your regular nursing bra. Ensure that the bra you
Ensure to turn
on the breast
pump before
inserting the
pump in your
bra.
To turn on your
breast pump,
press and hold
the power
button until the
lights switch on.
2.
Flashing white; pumping side to be selected
Solid white; side has been selected
The breast pump can
be used on the left
breast or the right
breast. Press the Left
selection button or the
Right selection button
to select which breast
you are going to
pump.