13. Troubleshooting

13.1 Vibrations and Notifications

One short vibration: When adjusting the vacuum by pressing the + button or the – button
Max/Min Vacuum has been reached.
Two short vibrations: The breast pump will pause automatically after 30 minutes. And also, if the
pumping session has not been restarted within 8 min of pausing.
18
Two short vibrations and milk status indicator flashing yellow: Check pumping position. Risk of
leaks. Move to upright position.
Two short vibrations and battery status indicator flashing red: Minimum charge level reached.
The breast pump will pause automatically. Finish the pumping session and charge the breast pump
afterwards.
Flashing blue; ready to be paired with the app
Solid blue; breast pump is connected to the app

13.2 Problems and Solutions

In case of an unexpected behavior of your breast pump check with the troubleshooting table below.
Find the issue in bold text, then follow the instructions listed below.
Problem
Power button not working
Solution
The breast pump does not switch on:
Press the power button for 1 second until the lights turn on.
• Ensure that the breast pump is not on the breast before switching on.
• Ensure that the breast pump is charged and the charging cable is not plugged
to a power source.
• Ensure that the breast pump is not overheated. Allow the breast pump to cool
down before switching it on.
If the breast pump still doesn't turn on, proceed with a factory reset by pressing
and holding the Power and the Start/Pause button for 10 seconds.
NOTE: a factory reset will erase the breast pump memory but all data stored
on the app will remain accessible.
The breast pump does not switch off:
The breast pump cannot be switched off during the pumping session. Press
Start/pause button to pause the session, break the latch with your fingers and
place the breast pump in a horizontal position until the breast pump turns off
automatically (could take up to 5 mins).