Herunterladen Inhalt Inhalt Diese Seite drucken

Connecting The Device; Connecting The Power Supply - Endress+Hauser RIA452 Betriebsanleitung

Vorschau ausblenden Andere Handbücher für RIA452:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

RIA452
"
"
!
Endress+Hauser
Terminal assignment
Terminal
Terminal assignment
11
+ 0/4 to 20 mA signal
12
Signal ground (current, voltage, temperature)
13
+ 1 V, + thermocouples, - resistance thermometer signal (3/4 wire)
15
+ resistance thermometer signal (4-wire)
17
+ 30 V
19
+ resistance thermometer power supply (3-wire/4-wire)
4. 2

Connecting the device

Caution!
Do not install or wire the device when it is connected to the power supply. Failure to comply with
this precaution can result in irreparable damage to the electronics.
4.2.1

Connecting the power supply

Caution!
• Before wiring the device, ensure that the supply voltage corresponds to the specification on the
nameplate.
• For the 90 to 250 V AC version (power supply connection), a switch marked as a separator, as
well as an overvoltage organ (rated current
device (easy to reach).
Fig. 6:
Connecting the power supply
4.2.2
Connecting external sensors
Note!
Active and passive sensors with analog, TC, resistance and RTD sensors can be attached to the
device.
Depending on the type of signal of the sensor in question, the terminals can be freely selected.
10 A), must be fitted in the supply line near the
Wiring
69

Quicklinks ausblenden:

Werbung

Kapitel

Inhaltsverzeichnis

Fehlerbehebung

loading

Inhaltsverzeichnis