17

Faults

If the fault cannot be eliminated with the help of the fault table, please inform Customer Service.
Please specify the type of centrifuge and the serial number. Both numbers can be found on the name plate of the
centrifuge.
Perform a MAINS RESET:
 Switch off the mains switch (switch position "0").
 Wait at least 10 seconds and then switch on the mains switch again (switch position "  ").
Fault
Display
No display
---
Tacho error
- 1 -
System reset
- 2 -
Communication
- 4 -
Overload
- 5 -
Overspeed
- 7 -
Lid error
- d -
Cause of fault
No voltage
Mains input fuses defective.
Failure of speed impulses during
operation.
Power failure during the centrifugation run.
(The centrifugation run was not finished.)
Fault in control unit or power unit.
Motor or motor control defective.
Fault in the power unit.
Error in lid locking or lid closure.
Remedy
Check distribution voltage.
-
Check mains power input fuse,
-
refer to Chapter "Change mains
input fuse".
Mains switch ON.
-
The device may not be switched off
-
as long as the rotation display
lit up and rotating. Wait until the
" " icon (lid closed) appears in the
rotation display (after about 100
seconds) and then run a "POWER
RESET".
When stationary, open lid and
-
press
key.
START
Repeat the centrifugation run if
-
necessary.
Perform a MAINS RESET when
-
the rotor has been stationary.
Perform a MAINS RESET when
-
the rotor has been stationary.
Perform a MAINS RESET when
-
the rotor has been stationary
Perform a MAINS RESET when
-
the rotor has been stationary.
EN
is
31/59

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis