3D System 100L, 200L, 500L Instructions for Use | Troubleshooting Guide
3.4 Emptying Flexsafe
Sartorius Stedim Biotech recommends dispensing through the bottom.
1. Connect bottom tubing line to receiving system according to standard
procedures for your site.
2. Open any clamps on the draining line (see Fig. 18). Drain the bag – gravity can
be used, but for faster draining and maximum product recovery,
Sartorius Stedim Biotech recommends the use of a peristaltic pump (see Fig. 24).
3. The bag collapses as it is drained. The sloped bottom of the Palletank
position of the drain hole provide nearly 100% fluid recovery. This efficiency can
be improved by lifting the bag at the end of the operation (see Fig. 25).
4. Once dispensing is completed, clamp the line (see Fig. 17) and disconnect the
Flexsafe
3D Bag from the receiving system.
®
5. Unlatch and open the bottom gate (see Fig. 7).
6. Lift the collapsed bag and remove it from the Palletank
7. Discard the Flexsafe
3D Bag in accordance with the procedures used
®
on your site.
4. Troubleshooting Guide
The bag will not fill.
Check that any clamps on the filling line are open and that the tubing
y
y
is not pinched.
When filling through the bottom, the bag comes out of the bottom gate.
The edge of the jaw is not closed inside the groove of the bottom port. Position it
y
y
correctly in the groove as described in step 10 of the installation procedure.
The bag looks very full before reaching the nominal volume:
The bag is most likely inserted backwards into the Palletank
If possible given the volume in the bag, open the bottom gate, spin the bag 180°
y
y
around the bottom gate and close the bottom gate. Examine Figures 15 & 16 for
proper bag positioning. It is critical that the bottom seam of the bag be along the
centerline of the Palletank
The bottom of the bag is incorrectly positioned.
Place the corners of the bag in the corners of the Palletank
y
y
10% of nominal capacity for 200L | 500L and 20% of nominal capacity for 100L is
filled before adjusting; otherwise bag failure may occur).
Many folds appear in the bag during filling.
The bag is too big for the Palletank
y
y
and use the bag suitable for the size of the Palletank
The bag does not reach the walls of the Palletank
The bag is too small for the Palletank
y
y
packaging and use the bag suitable for the size of the Palletank
The bag will not drain.
Check that any clamps on the drain line are open and that the tubing
y
y
is not pinched.
3D Bags 100L, 200L and 500L
®
for proper deployment.
®
, verify the bag size from the outer packaging
®
.
®
.
®
. Verify the bag size from the outer
®
and the
®
.
®
.
®
(make sure that only
®
.
®