3D System 100L, 200L, 500L Instructions for Use
3.3 Filling of 100L, 200L and 500L Flexsafe
Sartorius Stedim Biotech recommends filling through the top ports of bags.
1. Check that all clamps are closed (see Fig. 17).
2. Connect the bag to the supply line in accordance with the aseptic procedures
used on your site.
If you are using the Palletank
through the top, be particularly careful not to connect any tubing through
the front doors.
3. Open any clamps on the filling line (see Fig. 18).
4. Start to fill the Flexsafe
®
of peristaltic pumps for filling and dispensing from the bag (see Fig. 19).
– As the bag fills, make sure the liquid not get entrapped into the bag fold.
Regularly lift the bag from the top to have the liquid falling at the bottom of
the Palletank
(see Fig. 20)
®
– As the bag fills, make sure that no tubing becomes trapped between the bag
and the Palletank
®
for In-Process Fluid Handling and filling
®
3D Bag – Sartorius Stedim Biotech recommends the use
3D Bags
®