5. Commissioning, operation
5.4.2 Pin assignment with signal jump
Abbreviations, definitions
Signal Description
UB
Voltage source for sensor
UB+
Sensor-supply voltage (+)
Sensor-supply voltage (-)
UB-
UR
Voltage source for den signal jump
UR+
Signal jump-supply voltage (+)
UR-
Signal jump-supply voltage (-)
S+
Output signal (+)
S-
Output signal (-)
0V
0V-Potential
For models F5301 and F5308 with signal jump
Output 4 ... 20 mA, 2-wire
Circular connector M12 x 1, 4-pin
UB+/S+
4
+
UB
-
1
UB-
A
0V/S-
Output 4 ... 20 mA, 3-wire
Circular connector M12 x 1, 4-pin
UB+
S+
4
+
UB
-
1
A
0V/S-
Output 0...10 V, 3-wire
Circular connector M12 x 1, 4-pin
UB+
S+
4
+
-
UB
1
V
0V/S-
Cable colours only apply when using the WIKA standard cable, e.g. order no.: 14259454
Operating instructions, models F5301, F53C1, F5308, F53C8, F53S8
Signal
UR-
-
UB+/S+
UR
+
3
0V/S-
UR+
2
UR+
UR-
Shield
Signal
UB+
3
0V/S-
2
UR+
UR+
UR-
+
UR
-
S+
UR-
Shield
Signal
UB+
3
0V/S-
2
UR+
UR+
UR-
+
UR
-
S+
UR-
Shield
Signal
Description
Ammeter
A
Voltmeter
V
+
Voltage source
-
Switch
Shield [grounding]
4 ... 20 mA, 2-wire
1
3
2
4
Case / connector
4 ... 20 mA, 3-wire
1
3
2
3
4
Case / Connector
0 ... 10 V, 3-wire
1
3
2
3
4
Case / Connector
EN
Cable colour
Brown
Blue
White
Black
-
Cable colour
Brown
Blue
White
Blue
Black
-
Cable colour
Brown
Blue
White
Blue
Black
-
19