Error messages
The following list of error messages provides explanations of the error codes shown in the
display. When entering information, the error codes 0 to 9, A, B, C do not appear.
Error code
System errors
0
Pipeline not filled.
1
A/D converter
2
Positive or negative reference
too small
3
Flowrate greater than 130%
4
External zero return activated
5
RAM defective
1. Error 5 appears in the display:
second error 5 appears in Error
log only
7
Positive reference is too large
8
Negative reference is too large
6
Error > V
Error totalizer < R
Error totalizer
9
Excitation frequency is incorrect
A
MAX alarm limit value
B
MIN alarm limit value
C
Flowmeter primary data invalid
10
Entry > 1.00 Range DN > 10 m/s
11
Entry > 0.05 Range DN > 0.5 m/s Increase flow range.
16
Entry > 10% low flow cut-off
17
Entry < 0% low flow cut-off
20
Entry ≥ 100 s Damping
21
Entry < 0.5 s damping
22
Entry > 99 device address
38
Entry > 1000 pulse/unit
39
Entry < 0.001 pulse/unit
D184B133U02
Error removal
Open shut-off devices; fill pipeline; adjust Detector empty pipe
cut-off
Reduce flowrate, throttle shut-off device.
Check connection board and transmitter.
Reduce flowrate, change flow range.
Zero return activated by pump or field contact.
Program must be reinitialized.
Contact ABB Service department.
Information: Corrupted data in RAM, computer automatically
resets and uploads data from EEPROM.
Check signal cables and magnetic field excitation.
Check signal cables and magnetic field excitation.
Reset forward totalizer or preset new values in totalizer.
Reset reverse totalizer or preset new values in totalizer.
Forward/reverse or difference totalizer is defective
Reset forward/reverse totalizer
Check line frequency for supply power 50/60 Hz or for AC/DC
auxiliary power error in the digital signal board.
Reduce flowrate.
Increase flowrate.
The data for the flowmeter primary in the external EEPROM is
invalid. Compare data in the submenu "Flowmeter primary" with
data listed on the name plate. If the values match, use "Store
primary" to reset the error message. If the values are not
identical, the flowmeter primary data must first be reentered and
must then be completed with "Store primary". Contact ABB
Service.
Reduce flow .
Decrease input value.
Increase input value.
Decrease input value.
Increase entry value (as a function of the excitation frequency).
Decrease input value.
Decrease input value.
Increase input value.
FXE4000 (COPA-XE/MAG-XE)
Error messages
EN - 31