Herunterladen Diese Seite drucken

Agilent Technologies TPS-mini 9699180 Bedienungshandbuch Seite 285

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 29
Pressure gauge can be connected/disconnected from the TPS-mini
during normal operation.
Figure 18
The FRG700 gauge is intended for high vacuum use, so with pressure above 50
NOTE
mTorr the signal is misinterpreted and a "false" pressure is displayed. Use a
transition gauge to understand if the reading is really under 50 mTorr or not.
Pressure Display Module
Figure 19
TPS-mini User Manual / 87-901-003-01 (E)
Gauge Connector
Pressure display module
Technical Information
P1 – Vent
285/318
17

Werbung

loading

Diese Anleitung auch für:

Tps-mini 9699181Tps-mini 9699182