Herunterladen Diese Seite drucken

Agilent Technologies TPS-mini 9699180 Bedienungshandbuch Seite 262

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 29
16
Instructions for Use
Use
Use inert gas free from dust, particles or humidity (like Nitrogen) for venting the
CAUTION!
pump. The pressure at the vent port must be less than 2 bar (above atmospheric
pressure).
WARNING!
When employing the pump for pumping toxic, flammable, or radioactive
gases, please follow the required procedures for each gas disposal.
Do not use the pump in presence of explosive gases.
Switching on and Use of TPS-mini
To switch on the TPS-mini it is sufficient to supply the mains and
then move the external switch to ON position (red light on). The
integrated controller automatically recognizes the mains presence
and start up the pump.
At the first start up it is recommended to use the "Soft Start" mode by
enabling it on the controller. For the following start ups it is
recommended to disable the "Soft Start" mode. For the "Soft Start"
mode activation procedure, see the paragraph "Signal Description" in
the chapter "Technical Information".
262/318
TPS-mini User Manual / 87-901-003-01 (E)

Werbung

loading

Diese Anleitung auch für:

Tps-mini 9699181Tps-mini 9699182