Operation
Acknowledge pre-alarm:
Press the [OK] key, only the audible alarm and the vibration alarm are
switched off.
The battery still lasts approx. 20 minutes after the first battery pre-alarm.
6.6.5

Battery main alarm

Intermittent alarm:
Flashing special symbol »
display.
The battery main alarm cannot be acknowledged or cancelled.
The instrument is automatically switched off again after 10 seconds.
Before the instrument is switched off, the visual, audible and vibration
alarms are activated for a short time.
6.6.6
Instrument alarm
Intermittent alarm:
Special symbol »
the display:
The instrument is not ready for operation.
Contact maintenance or Draeger Service to rectify the problem.
6.7

Info Mode

6.7.1
Activating the Info mode
In measuring mode, press the [OK] key for approx. 3 seconds.
If any warning or fault messages exist, the corresponding note or error
codes are displayed (see Technical Handbook).
Press the [OK] key successively for the next display. The peak values and
the exposure values TWA and STEV will be displayed.
If no key is pressed for 10 seconds, the instrument returns automatically to
measuring mode.
6.7.2
Info Off mode
Press the [ + ] key when the instrument is turned off.
The name of the gas, measuring unit, and measuring range limit value are
displayed for all channels.
Press the [ + ] key again to exit the Info Off Mode (or via timeout).
6.7.3
Calling the Quick Menu
In measuring mode, press the [ + ] key three times.
40
« on the right side of the
« displayed on the right side of