Operation
Varta Type 4006
Alkaline batteries – T4 – (non rechargeable!)
Duracell Procell MN1500
NiMH rechargeable batteries – T3 – (rechargeable)
GP 180AAHC
Only charge NiMH power packs T4 (type HBT 0000) or T4 HC (type
HBT 0100) with the appropriate Dräger charger. Charge NiMH single cells
for ABT 0100 battery holder as directed by the manufacturer. Ambient
temperature during charging: 0 to +40 °C.
1.
Switching off the instrument: Press and hold the [OK] and [ + ] keys
simultaneously.
2.
Loosen the screw on the power pack and remove the power pack.
Battery holder (order no. 83 22 237): Replace alkaline batteries or NiMHy
rechargeable batteries. Ensure correct polarity.
!
WARNING
Explosion hazard!
The gas detector must only be operated with the ABT 0100 battery holder.
NiMH power pack T4 (Type HBT 0000) / T4 HC (Type HBT 0100):
3.
replace complete power pack.
4.
Insert the power pack into the instrument and tighten the screw, the
instrument switches on automatically.
6.8.1
Charge instrument with NiMH power pack T4 (Type HBT 0000)/ T4
HC (Type HBT 0100)
!
WARNING
Explosion hazard! To reduce the risk of ignition of a flammable or explosive
atmosphere, strictly adhere to the following warning statements:
Do not charge underground or in explosion hazard areas! The chargers are not
designed in accordance with the regulations for fire damp and explosion protection.
Charge NiMH power packs T4 (type HBT 0000) or T4 HC (type HBT 0100) with
the appropriate Dräger charger. Ambient temperature during charging: 0 to +40 °C.
1)
Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.
42
1)
(industrial)
1)
1)
(1800 mAh) max. 40 °C ambient temperature.