Even if the instrument is not used, we recommend that you store the instrument
in the charger!
Insert the switched off instrument into the charger module.
Display LED on the charger module:
To protect the battery charge only in the temperature range of 5 to 35 °C. Outside
this temperature range, the charging process is automatically interrupted and
automatically continued after the temperature range has been reached again.
The charging time is typically 4 hours. A new NiMH power pack reaches its full
capacity after three complete charge/discharge cycles. Never store the
instrument for extended periods without being connected to a power source
(maximum of 2 months) because the internal buffer battery will drain.
6.9
Carrying out the manual bump test
In the case of a manual function test the effect of the H
taken into account accordingly.
A potentially activated H
during a manual calibration, a PC calibration or an automatic Bump Test for
the relevant duration.
The automatic functional test with the Bump Test Station is described in the
instructions for use for the bump test station and in the Technical Manual.
1.
Prepare a test gas cylinder, the volume flow must be 0.5 L/min and the gas
concentration must be higher than the alarm threshold concentration that is
to be tested.
2.
Connect the test gas cylinder with the calibration cradle (order no. 83 18 752).
!
CAUTION
Risk to health! Never inhale the test gas.
Observe the hazard warnings of the relevant Safety Data Sheets.
Charge
Malfunction
Full
offsetting is automatically temporarily deactivated
2
Operation
en
offsetting must be
2
43